Comparative study of symmetric and asymmetric cylindrical MRI gradient Y-coils in terms of induced eddy currents
Contribuinte(s) |
Y.T. Zhang |
---|---|
Data(s) |
01/01/2005
|
Resumo |
We have recently introduced the concept of whole-body asymmetric MRI systems [1]. In this theoretical study, we investigate the PNS characteristics of whole-body asymmetric gradient systems as compared to conventional symmetric systems. Recent experimental evidence [2] supports the hypothesis of transverse gradients being the largest contributor of PNS due to induced electric currents. Asymmetric head gradient coils have demonstrated benefits in the past [3]. The numerical results are based on an anatomically-accurate 2mm-human voxel-phantom NORMAN [4]. The results of this study can facilitate the optimization of whole-body asymmetric gradients in terms of patient comfort/safety (less PNS), while prospering the use of asymmetric MRI systems for in-vivo medical interventions. |
Identificador | |
Publicador |
IEEE |
Palavras-Chave | #E1 #291500 Biomedical Engineering #729901 Technological and organisational innovation |
Tipo |
Conference Paper |