Numerical modeling of 11.1 T MRI of a human head using MoM/FDTD methods


Autoria(s): Liu, Feng; Beck, Barbara L.; Xu, Bin; Fitzsimmons, Jeffrey R.; Blackband, Stephen J.; Crozier, Stuart
Contribuinte(s)

D. D. Traficante

Data(s)

01/02/2005

Resumo

In a recent study, severe distortions in the proton images of an excised, fixed, human brain in an 11.1 Tesla/40 cm MR instrument have been observed, and the effect modeled on phantom images using a finite difference time domain (FDTD) model. in the present study, we extend these simulations to that of a complete human head, employing a hybrid FDTD and method of moments (MoM) approach, which provides a validated method for simulating biological samples in coil structures. The effect of fixative on the image distortions is explored. importantly, temperature distributions within the head are also simulated using a bioheat method based on parameters derived from the electromagnetic simulations. The MoM/FDTD simulations confirm that the transverse magnetic field (B,) from a ReCav resonator exhibits good homogeneity in air but strong inhomogeneity when loaded with the head with or without fixative. The fixative serves to increase the distortions, but they are still significant for the in vivo simulations. The simulated signal intensity (SI) distribution within the sample confirm the distortions in the experimental images are caused by the complex interactions of the incident electromagnetic fields with tissue, which is heterogeneous in terms of conductivity and permittivity. The temperature distribution is likewise heterogeneous, raising concerns regarding hot spot generation in the sample that may exceed acceptable levels in future in vivo studies. As human imaging at 11.1 T is some time away, simulations are important in terms of predicting potential safety issues as well as evaluating practical concerns about the quality of images. Simulation on a whole human head at 11.1 T implies the wave behavior presents significant engineering challenges for ultra-high-field (UHF) MRI. Novel strategies will have to be employed in imaging technique and resonator design for UHF MRI to achieve the theoretical signal-to-noise ratio (SNR) improvements it offers over lower field systems. (C) 2005 Wiley Periodicals, Inc.

Identificador

http://espace.library.uq.edu.au/view/UQ:77883

Idioma(s)

eng

Publicador

John Wiley & Sons

Palavras-Chave #Instruments & Instrumentation #Physics, Atomic, Molecular & Chemical #Spectroscopy #Ultra-high-field Mri #291500 Biomedical Engineering #729901 Technological and organisational innovation
Tipo

Journal Article