A novel scanning lens instrument for evaluating Fresnel lens performance: equipment development and initial results
Data(s) |
2013
|
---|---|
Resumo |
A system dedicated to the optical transmittance characterization of Fresnel lenses has been developed at NREL, in collaboration with the UPM. The system quantifies the optical efficiency of the lens by generating a performance map. The shape of the focused spot may also be analyzed to understand change in the lens performance. The primary instrument components (lasers and CCD detector) have been characterized to confirm their capability for performing optical transmittance measurements. Measurements performed on SoG and PMMA lenses subject to a variety of indoor conditions (e.g., UV and damp heat) identified differences in the optical efficiency of the evaluated lenses, demonstrating the ability of the Scanning Lens Instrument (SLI) to distinguish between the aged lenses. |
Formato |
application/pdf |
Identificador | |
Idioma(s) |
eng |
Relação |
http://oa.upm.es/31087/1/INVE_MEM_2013_175659.pdf http://dx.doi.org/10.1063/1.4822213 info:eu-repo/semantics/altIdentifier/doi/10.1063/1.4822213 |
Direitos |
http://creativecommons.org/licenses/by-nc-nd/3.0/es/ info:eu-repo/semantics/openAccess |
Fonte |
9th International Conference on Concentrator Photovoltaic Systems | 9th International Conference on Concentrator Photovoltaic Systems | 05/04/2013 – 17/04/2013 | Miyazaki, Japan |
Palavras-Chave | #Telecomunicaciones |
Tipo |
info:eu-repo/semantics/conferenceObject Ponencia en Congreso o Jornada PeerReviewed |