A novel scanning lens instrument for evaluating Fresnel lens performance: equipment development and initial results


Autoria(s): Herrero Martin, Rebeca; Miller, David C.; Kurtz, Sarah R.; Anton Hernandez, Ignacio; Sala Pano, Gabriel
Data(s)

2013

Resumo

A system dedicated to the optical transmittance characterization of Fresnel lenses has been developed at NREL, in collaboration with the UPM. The system quantifies the optical efficiency of the lens by generating a performance map. The shape of the focused spot may also be analyzed to understand change in the lens performance. The primary instrument components (lasers and CCD detector) have been characterized to confirm their capability for performing optical transmittance measurements. Measurements performed on SoG and PMMA lenses subject to a variety of indoor conditions (e.g., UV and damp heat) identified differences in the optical efficiency of the evaluated lenses, demonstrating the ability of the Scanning Lens Instrument (SLI) to distinguish between the aged lenses.

Formato

application/pdf

Identificador

http://oa.upm.es/31087/

Idioma(s)

eng

Relação

http://oa.upm.es/31087/1/INVE_MEM_2013_175659.pdf

http://dx.doi.org/10.1063/1.4822213

info:eu-repo/semantics/altIdentifier/doi/10.1063/1.4822213

Direitos

http://creativecommons.org/licenses/by-nc-nd/3.0/es/

info:eu-repo/semantics/openAccess

Fonte

9th International Conference on Concentrator Photovoltaic Systems | 9th International Conference on Concentrator Photovoltaic Systems | 05/04/2013 – 17/04/2013 | Miyazaki, Japan

Palavras-Chave #Telecomunicaciones
Tipo

info:eu-repo/semantics/conferenceObject

Ponencia en Congreso o Jornada

PeerReviewed