Proof of concept of an imaging system demonstrator for PET applications with SiPM
Data(s) |
01/08/2013
|
---|---|
Resumo |
A PET imaging system demonstrator based on LYSO crystal arrays coupled to SiPM matrices is under construction at the University and INFN of Pisa. Two SiPM matrices, composed of 8×8 SiPM pixels, and 1,5 mm pitch, have been coupled one to one to a LYSO crystals array and read out by a custom electronics system. front-end ASICs were used to read 8 channels of each matrix. Data from each front-end were multiplexed and sent to a DAQ board for the digital conversion; a motherboard collects the data and communicates with a host computer through a USB port for the storage and off-line data processing. In this paper we show the first preliminary tomographic image of a point-like radioactive source acquired with part of the two detection heads in time coincidence. |
Formato |
application/pdf |
Identificador | |
Idioma(s) |
eng |
Publicador |
E.T.S.I. Telecomunicación (UPM) |
Relação |
http://oa.upm.es/28895/1/INVE_MEM_2013_166151.pdf http://www.sciencedirect.com/science/article/pii/S0168900212015665 info:eu-repo/semantics/altIdentifier/doi/10.1016/j.nima.2012.12.033 |
Direitos |
http://creativecommons.org/licenses/by-nc-nd/3.0/es/ info:eu-repo/semantics/openAccess |
Fonte |
Nuclear Instruments and Methods in Physics Research A: : Accelerators, Spectrometers, Detectors and Associated Equipment, ISSN 0168-9002, 2013-08, Vol. 720 |
Palavras-Chave | #Telecomunicaciones |
Tipo |
info:eu-repo/semantics/article Artículo PeerReviewed |