Low-resource synchronous coincidence processor for positron emission tomography


Autoria(s): Sportelli, Giancarlo; Belcari, Nicola; Guerra Gutiérrez, Pedro; Santos Lleo, Andres de
Data(s)

01/08/2011

Resumo

We developed a new FPGA-based method for coincidence detection in positronemissiontomography. The method requires low device resources and no specific peripherals in order to resolve coincident digital pulses within a time window of a few nanoseconds. This method has been validated with a low-end Xilinx Spartan-3E and provided coincidence resolutions lower than 6 ns. This resolution depends directly on the signal propagation properties of the target device and the maximum available clock frequency, therefore it is expected to improve considerably on higher-end FPGAs.

Formato

application/pdf

Identificador

http://oa.upm.es/12116/

Idioma(s)

eng

Publicador

E.T.S.I. Telecomunicación (UPM)

Relação

http://oa.upm.es/12116/2/INVE_MEM_2011_104080.pdf

http://www.sciencedirect.com/science/article/pii/S0168900210027178

info:eu-repo/semantics/altIdentifier/doi/10.1016/j.nima.2010.11.170

Direitos

http://creativecommons.org/licenses/by-nc-nd/3.0/es/

info:eu-repo/semantics/openAccess

Fonte

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, ISSN 0168-9002, 2011-08, Vol. 648, No. Supple