Bilateral deep brain stimulation: the placement of the second electrode is not necessarily less accurate than that of the first one.


Autoria(s): Sadeghi, Yalda; Pralong, Etienne; Knebel, Jean-François; Vingerhoets, François; Pollo, Claudio; Levivier, Marc; Bloch, Jocelyne
Data(s)

2015

Resumo

BACKGROUND Deep brain stimulation (DBS) is recognized as an effective treatment for movement disorders. We recently changed our technique, limiting the number of brain penetrations to three per side. OBJECTIVES The first aim was to evaluate the electrode precision on both sides of surgery since we implemented this surgical technique. The second aim was to analyse whether or not the electrode placement was improved with microrecording and macrostimulation. METHODS We retrospectively reviewed operation protocols and MRIs of 30 patients who underwent bilateral DBS. For microrecording and macrostimulation, we used three parallel channels of the 'Ben Gun' centred on the MRI-planned target. Pre- and post-operative MRIs were merged. The distance between the planned target and the centre of the implanted electrode artefact was measured. RESULTS There was no significant difference in targeting precision on both sides of surgery. There was more intra-operative adjustment of the second electrode positioning based on microrecording and macrostimulation, which allowed to significantly approach the MRI-planned target on the medial-lateral axis. CONCLUSION There was more electrode adjustment needed on the second side, possibly in relation with brain shift. We thus suggest performing a single central track with electrophysiological and clinical assessment, with multidirectional exploration on demand for suboptimal clinical responses.

Formato

application/pdf

Identificador

http://boris.unibe.ch/80158/1/368439.pdf

Sadeghi, Yalda; Pralong, Etienne; Knebel, Jean-François; Vingerhoets, François; Pollo, Claudio; Levivier, Marc; Bloch, Jocelyne (2015). Bilateral deep brain stimulation: the placement of the second electrode is not necessarily less accurate than that of the first one. Stereotactic and functional neurosurgery, 93(3), pp. 160-167. Karger 10.1159/000368439 <http://dx.doi.org/10.1159/000368439>

doi:10.7892/boris.80158

info:doi:10.1159/000368439

info:pmid:25791181

urn:issn:1011-6125

Idioma(s)

eng

Publicador

Karger

Relação

http://boris.unibe.ch/80158/

Direitos

info:eu-repo/semantics/openAccess

Fonte

Sadeghi, Yalda; Pralong, Etienne; Knebel, Jean-François; Vingerhoets, François; Pollo, Claudio; Levivier, Marc; Bloch, Jocelyne (2015). Bilateral deep brain stimulation: the placement of the second electrode is not necessarily less accurate than that of the first one. Stereotactic and functional neurosurgery, 93(3), pp. 160-167. Karger 10.1159/000368439 <http://dx.doi.org/10.1159/000368439>

Palavras-Chave #610 Medicine & health
Tipo

info:eu-repo/semantics/article

info:eu-repo/semantics/publishedVersion

PeerReviewed