Influence of Abutment Design on Stiffness, Strength, and Failure of Implant-Supported Monolithic Resin Nano Ceramic (RNC) Crowns
Data(s) |
01/12/2015
|
---|---|
Resumo |
BACKGROUND Recent technical development allows the digital manufacturing of monolithic reconstructions with high-performance materials. For implant-supported crowns, the fixation requires an abutment design onto which the reconstruction can be bonded. PURPOSE The aim of this laboratory investigation was to analyze stiffness, strength, and failure modes of implant-supported, computer-assisted design and computer-aided manufacturing (CAD/CAM)-generated resin nano ceramic (RNC) crowns bonded to three different titanium abutments. MATERIALS AND METHODS Eighteen monolithic RNC crowns were produced and loaded in a universal testing machine under quasi-static condition according to DIN ISO 14801. With regard to the type of titanium abutment, three groups were defined: (1) prefabricated cementable standard; (2) CAD/CAM-constructed individualized; and (3) novel prefabricated bonding base. Stiffness and strength were measured and analyzed statistically with Wilcoxon rank sum test. Sections of the specimens were examined microscopically. RESULTS Stiffness demonstrated high stability for all specimens loaded in the physiological loading range with means and standard deviations of 1,579 ± 120 N/mm (group A), 1,733 ± 89 N/mm (group B), and 1,704 ± 162 N/mm (group C). Mean strength of the novel prefabricated bonding base (group C) was 17% lower than of the two other groups. Plastic deformations were detectable for all implant-abutment crown connections. CONCLUSIONS Monolithic implant crowns made of RNC seem to represent a feasible and stable prosthetic construction under laboratory testing conditions with strength higher than the average occlusal force, independent of the different abutment designs used in this investigation. |
Formato |
application/pdf |
Identificador |
http://boris.unibe.ch/77417/1/In%EF%AC%82uence%20of%20Abutment%20Design.pdf Joda, Tim; Huber, Samuel; Bürki, Alexander; Zysset, Philippe; Brägger, Urs (2015). Influence of Abutment Design on Stiffness, Strength, and Failure of Implant-Supported Monolithic Resin Nano Ceramic (RNC) Crowns. Clinical implant dentistry and related research, 17(6), pp. 1200-1207. Blackwell 10.1111/cid.12215 <http://dx.doi.org/10.1111/cid.12215> doi:10.7892/boris.77417 info:doi:10.1111/cid.12215 info:pmid:24629091 urn:issn:1523-0899 |
Idioma(s) |
eng |
Publicador |
Blackwell |
Relação |
http://boris.unibe.ch/77417/ |
Direitos |
info:eu-repo/semantics/restrictedAccess |
Fonte |
Joda, Tim; Huber, Samuel; Bürki, Alexander; Zysset, Philippe; Brägger, Urs (2015). Influence of Abutment Design on Stiffness, Strength, and Failure of Implant-Supported Monolithic Resin Nano Ceramic (RNC) Crowns. Clinical implant dentistry and related research, 17(6), pp. 1200-1207. Blackwell 10.1111/cid.12215 <http://dx.doi.org/10.1111/cid.12215> |
Palavras-Chave | #570 Life sciences; biology #610 Medicine & health |
Tipo |