End caps prevent nail migration in elastic stable intramedullary nailing in paediatric femoral fractures: a biomechanical study using synthetic and cadaveric bones.


Autoria(s): Windolf, M; Fischer, M F; Popp, Albrecht; Matthys, R; Schwieger, K; Gueorguiev, B; Hunter, J B; Slongo, Theddy
Data(s)

01/04/2015

Resumo

End caps are intended to prevent nail migration (push-out) in elastic stable intramedullary nailing. The aim of this study was to investigate the force at failure with and without end caps, and whether different insertion angles of nails and end caps would alter that force at failure. Simulated oblique fractures of the diaphysis were created in 15 artificial paediatric femurs. Titanium Elastic Nails with end caps were inserted at angles of 45°, 55° and 65° in five specimens for each angle to create three study groups. Biomechanical testing was performed with axial compression until failure. An identical fracture was created in four small adult cadaveric femurs harvested from two donors (both female, aged 81 and 85 years, height 149 cm and 156 cm, respectively). All femurs were tested without and subsequently with end caps inserted at 45°. In the artificial femurs, maximum force was not significantly different between the three groups (p = 0.613). Push-out force was significantly higher in the cadaveric specimens with the use of end caps by an up to sixfold load increase (830 N, standard deviation (SD) 280 vs 150 N, SD 120, respectively; p = 0.007). These results indicate that the nail and end cap insertion angle can be varied within 20° without altering construct stability and that the risk of elastic stable intramedullary nailing push-out can be effectively reduced by the use of end caps.

Formato

application/pdf

Identificador

http://boris.unibe.ch/69220/1/Artikel%202015%20Slongo%201.pdf

Windolf, M; Fischer, M F; Popp, Albrecht; Matthys, R; Schwieger, K; Gueorguiev, B; Hunter, J B; Slongo, Theddy (2015). End caps prevent nail migration in elastic stable intramedullary nailing in paediatric femoral fractures: a biomechanical study using synthetic and cadaveric bones. The Bone & Joint Journal, 97-B(4), pp. 558-563. British Editorial Society of Bone and Joint Surgery 10.1302/0301-620X.97B4.34169 <http://dx.doi.org/10.1302/0301-620X.97B4.34169>

doi:10.7892/boris.69220

info:doi:10.1302/0301-620X.97B4.34169

info:pmid:25820898

urn:issn:2049-4394

Idioma(s)

eng

Publicador

British Editorial Society of Bone and Joint Surgery

Relação

http://boris.unibe.ch/69220/

Direitos

info:eu-repo/semantics/restrictedAccess

Fonte

Windolf, M; Fischer, M F; Popp, Albrecht; Matthys, R; Schwieger, K; Gueorguiev, B; Hunter, J B; Slongo, Theddy (2015). End caps prevent nail migration in elastic stable intramedullary nailing in paediatric femoral fractures: a biomechanical study using synthetic and cadaveric bones. The Bone & Joint Journal, 97-B(4), pp. 558-563. British Editorial Society of Bone and Joint Surgery 10.1302/0301-620X.97B4.34169 <http://dx.doi.org/10.1302/0301-620X.97B4.34169>

Palavras-Chave #610 Medicine & health
Tipo

info:eu-repo/semantics/article

info:eu-repo/semantics/publishedVersion

PeerReviewed