Distal interlocking screw placement in the femur: free-hand versus electromagnetic assisted technique (sureshot).


Autoria(s): Maqungo, Sithombo; Horn, Anria; Bernstein, Brian; Keel, Marius; Roche, Stephen
Data(s)

01/12/2014

Resumo

OBJECTIVES To compare the free-hand (FH) technique of placing interlocking screws to a commercially available electromagnetic (EM) targeting system in terms of operating time, radiation dose, and accuracy of screw placement. METHODS Between September 2011 and July 2012, we prospectively randomized 100 consecutive femur shaft fractures in 99 patients requiring intramedullary nails to either FH using fluoroscopy (n = 43) or EM targeting (n = 38; Sureshot). SETTING Single Level 1 University Hospital Trauma Center. MAIN OUTCOME MEASUREMENTS The 2 groups were assessed for distal locking with respect to time, radiation, and accuracy. RESULTS Eight-one fractures had data accurately recorded (38 EM/43 FH). The average total operative time was 50 minutes (range, 25-88 minutes; SD, 13.9 minutes) for the FH group and 57 minutes (range, 40-103 minutes; SD, 16.12 minutes) for the EM group. The average time for distal locking was 10 minutes (range, 4-16 minutes; SD, 3.56 minutes) with FH and 11 minutes (range, 6-28 minutes; SD, 10.24 minutes) with EM. Average radiation dose for distal locking was significantly less (P < 0.0001) for EM at 230.54 μGy (range, 51-660 μGy; SD, 0.17 μGy) compared with 690.27 μGy (range, 200-2310 μGy; SD, 0.52 μGy) for FH. There were 2 misplaced drill bits in FH and 3 in EM. This was not statistically significant (P = 0.888). CONCLUSIONS The electromagnetic targeting device (Sureshot) significantly reduced radiation exposure during placement of distal interlocking screws, without sacrificing operative time, and was equivalent in accuracy when compared with the FH technique. LEVEL OF EVIDENCE Therapeutic level II.

Formato

application/pdf

Identificador

http://boris.unibe.ch/62963/1/00005131-201412000-00013.pdf

Maqungo, Sithombo; Horn, Anria; Bernstein, Brian; Keel, Marius; Roche, Stephen (2014). Distal interlocking screw placement in the femur: free-hand versus electromagnetic assisted technique (sureshot). Journal of orthopaedic trauma, 28(12), e281-e283. Lippincott Williams & Wilkins 10.1097/BOT.0000000000000125 <http://dx.doi.org/10.1097/BOT.0000000000000125>

doi:10.7892/boris.62963

info:doi:10.1097/BOT.0000000000000125

info:pmid:24714403

urn:issn:0890-5339

Idioma(s)

eng

Publicador

Lippincott Williams & Wilkins

Relação

http://boris.unibe.ch/62963/

Direitos

info:eu-repo/semantics/openAccess

Fonte

Maqungo, Sithombo; Horn, Anria; Bernstein, Brian; Keel, Marius; Roche, Stephen (2014). Distal interlocking screw placement in the femur: free-hand versus electromagnetic assisted technique (sureshot). Journal of orthopaedic trauma, 28(12), e281-e283. Lippincott Williams & Wilkins 10.1097/BOT.0000000000000125 <http://dx.doi.org/10.1097/BOT.0000000000000125>

Palavras-Chave #610 Medicine & health
Tipo

info:eu-repo/semantics/article

info:eu-repo/semantics/publishedVersion

PeerReviewed