Comparison of the biomechanical properties of a ventral cervical intervertebral anchored fusion device with locking plate fixation applied to cadaveric canine cervical spines


Autoria(s): Schöllhorn, Bernhard; Bürki, Alexander; Stahl, Christina; Howard, J.; Forterre, Franck
Data(s)

07/09/2013

Resumo

OBJECTIVE: To evaluate fixation properties of a new intervertebral anchored fusion device and compare these with ventral locking plate fixation. STUDY DESIGN: In vitro biomechanical evaluation. ANIMALS: Cadaveric canine C4-C7 cervical spines (n = 9). METHODS: Cervical spines were nondestructively loaded with pure moments in a nonconstraining testing apparatus to induce flexion/extension while angular motion was measured. Range of motion (ROM) and neutral zone (NZ) were calculated for (1) intact specimens, (2) specimens after discectomy and fixation with a purpose-built intervertebral fusion cage with integrated ventral fixation, and (3) after removal of the device and fixation with a ventral locking plate. RESULTS: Both fixation techniques resulted in a decrease in ROM and NZ (P < .001) compared with the intact segments. There were no significant differences between the anchored spacer and locking plate fixation. CONCLUSION: An anchored spacer appears to provide similar biomechanical stability to that of locking plate fixation.

Formato

application/pdf

Identificador

http://boris.unibe.ch/40380/1/vsu12044.pdf

Schöllhorn, Bernhard; Bürki, Alexander; Stahl, Christina; Howard, J.; Forterre, Franck (2013). Comparison of the biomechanical properties of a ventral cervical intervertebral anchored fusion device with locking plate fixation applied to cadaveric canine cervical spines. Veterinary surgery, 42(7), pp. 825-831. Wiley-Blackwell 10.1111/j.1532-950X.2013.12044.x <http://dx.doi.org/10.1111/j.1532-950X.2013.12044.x>

doi:10.7892/boris.40380

info:doi:10.1111/j.1532-950X.2013.12044.x

info:pmid:24033669

urn:issn:0161-3499

Idioma(s)

eng

Publicador

Wiley-Blackwell

Relação

http://boris.unibe.ch/40380/

http://www.ncbi.nlm.nih.gov/pubmed/24033669

Direitos

info:eu-repo/semantics/restrictedAccess

Fonte

Schöllhorn, Bernhard; Bürki, Alexander; Stahl, Christina; Howard, J.; Forterre, Franck (2013). Comparison of the biomechanical properties of a ventral cervical intervertebral anchored fusion device with locking plate fixation applied to cadaveric canine cervical spines. Veterinary surgery, 42(7), pp. 825-831. Wiley-Blackwell 10.1111/j.1532-950X.2013.12044.x <http://dx.doi.org/10.1111/j.1532-950X.2013.12044.x>

Palavras-Chave #610 Medicine & health
Tipo

info:eu-repo/semantics/article

info:eu-repo/semantics/publishedVersion

PeerReviewed