Cone Beam and Micro-Computed Tomography Validation of Manual Array Insertion for Minimally Invasive Cochlear Implantation


Autoria(s): Wimmer, Wilhelm; Bell, Brett; Huth, Markus; Weisstanner, Christian; Gerber, Nicolas; Kompis, Martin; Weber, Stefan; Caversaccio, Marco
Data(s)

2014

Resumo

Delivering cochlear implants through a minimally invasive tunnel (1.8 mm in diameter) from the mastoid surface to the inner ear is referred to as direct cochlear access (DCA). Based on cone beam as well as micro-computed tomography imaging, this in vitro study evaluates the feasibility and efficacy of manual cochlear electrode array insertions via DCA. Free-fitting electrode arrays were inserted in 8 temporal bone specimens with previously drilled DCA tunnels. The insertion depth angle, procedural time, tunnel alignment as well as the inserted scala and intracochlear trauma were assessed. Seven of the 8 insertions were full insertions, with insertion depth angles higher than 520°. Three cases of atraumatic scala tympani insertion, 3 cases of probable basilar membrane rupture and 1 case of dislocation into the scala vestibuli were observed (1 specimen was damaged during extraction). Manual electrode array insertion following a DCA procedure seems to be feasible and safe and is a further step toward clinical application of image-guided otological microsurgery.

Formato

application/pdf

Identificador

http://boris.unibe.ch/39901/1/356165.pdf

Wimmer, Wilhelm; Bell, Brett; Huth, Markus; Weisstanner, Christian; Gerber, Nicolas; Kompis, Martin; Weber, Stefan; Caversaccio, Marco (2014). Cone Beam and Micro-Computed Tomography Validation of Manual Array Insertion for Minimally Invasive Cochlear Implantation. Audiology & Neurotology, 19(1), pp. 22-30. Karger 10.1159/000356165 <http://dx.doi.org/10.1159/000356165>

doi:10.7892/boris.39901

info:doi:10.1159/000356165

info:pmid:24280962

urn:issn:1420-3030, 1421-9700

Idioma(s)

eng

Publicador

Karger

Relação

http://boris.unibe.ch/39901/

Direitos

info:eu-repo/semantics/restrictedAccess

Fonte

Wimmer, Wilhelm; Bell, Brett; Huth, Markus; Weisstanner, Christian; Gerber, Nicolas; Kompis, Martin; Weber, Stefan; Caversaccio, Marco (2014). Cone Beam and Micro-Computed Tomography Validation of Manual Array Insertion for Minimally Invasive Cochlear Implantation. Audiology & Neurotology, 19(1), pp. 22-30. Karger 10.1159/000356165 <http://dx.doi.org/10.1159/000356165>

Palavras-Chave #610 Medicine & health #570 Life sciences; biology
Tipo

info:eu-repo/semantics/article

info:eu-repo/semantics/publishedVersion