An examination chair to measure internal rotation of the hip in routine settings: a validation study


Autoria(s): Reichenbach, Stephan; Jüni, Peter; Nüesch, Eveline; Frey, Franz; Ganz, Reinhold; Leunig, Michael
Data(s)

2010

Resumo

OBJECTIVE: To determine the performance of a newly developed examination chair as compared with the clinical standard of assessing internal rotation (IR) of the flexed hip with a goniometer. METHODS: The examination chair allowed measurement of IR in a sitting position simultaneously in both hips, with hips and knees flexed 90 degrees, lower legs hanging unsupported and a standardized load of 5 kg applied to both ankles using a bilateral pulley system. Clinical assessment of IR was performed in supine position with hips and knees flexed 90 degrees using a goniometer. Within the framework of a population-based inception cohort study, we calculated inter-observer agreement in two samples of 84 and 64 consecutive, unselected young asymptomatic males using intra-class correlation coefficients (ICC) and determined the correlation between IR assessed with examination chair and clinical assessment. RESULTS: Inter-observer agreement was excellent for the examination chair (ICC right hip, 0.92, 95% confidence interval [CI] 0.89-0.95; ICC left hip, 0.90, 95% CI 0.86-0.94), and considerably higher than that seen with clinical assessment (ICC right hip, 0.65, 95% CI 0.49-0.77; ICC left hip, 0.69, 95% CI 0.54-0.80, P for difference in ICC between examination chair and clinical assessment <or=0.001). The average range of motion (ROM) obtained with examination chair and clinical assessment were similar (difference 1.1 degrees, 95% CI - 0.7-2.8 degrees, P=0.23), and the correlation was strong (Pearson's coefficient, 0.75, 95% CI 0.62-0.84). CONCLUSIONS: The use of the examination chair resulted in a precise assessment of hip IR in our population-based inception cohort study of young asymptomatic males. It was strongly correlated with standard clinical assessment of IR but was considerably more reliable.

Formato

application/pdf

Identificador

http://boris.unibe.ch/43/1/43.pdf

Reichenbach, Stephan; Jüni, Peter; Nüesch, Eveline; Frey, Franz; Ganz, Reinhold; Leunig, Michael (2010). An examination chair to measure internal rotation of the hip in routine settings: a validation study. Osteoarthritis and cartilage, 18(3), pp. 365-371. Amsterdam: Elsevier 10.1016/j.joca.2009.10.001 <http://dx.doi.org/10.1016/j.joca.2009.10.001>

doi:10.7892/boris.43

info:doi:10.1016/j.joca.2009.10.001

info:pmid:19833251

urn:issn:1063-4584

Idioma(s)

eng

Publicador

Elsevier

Relação

http://boris.unibe.ch/43/

Direitos

info:eu-repo/semantics/openAccess

Fonte

Reichenbach, Stephan; Jüni, Peter; Nüesch, Eveline; Frey, Franz; Ganz, Reinhold; Leunig, Michael (2010). An examination chair to measure internal rotation of the hip in routine settings: a validation study. Osteoarthritis and cartilage, 18(3), pp. 365-371. Amsterdam: Elsevier 10.1016/j.joca.2009.10.001 <http://dx.doi.org/10.1016/j.joca.2009.10.001>

Palavras-Chave #610 Medicine & health #360 Social problems & social services
Tipo

info:eu-repo/semantics/article

info:eu-repo/semantics/publishedVersion

PeerReviewed