Monte Carlo Semi-Empirical Model for Si(Li) X-Ray Detector: Differences between Nominal and Fitted Parameters.


Autoria(s): López-Pino, N.; Padilla-Cabal, F.; Alvarez, Juan Alejandro Garcia; Vázquez, L.; Alessandro, K. D.; Alfonso, Camilo Miguel Correa; Godoy, Walquiria; Maidana, Nora Lia; Vanin, Vito Roberto
Contribuinte(s)

UNIVERSIDADE DE SÃO PAULO

Data(s)

16/04/2014

16/04/2014

02/09/2012

Resumo

A detailed characterization of a X-ray Si(Li) detector was performed to obtain the energy dependence of efficiency in the photon energy range of 6.4 - 59.5 keV. which was measured and reproduced by Monte Carlo (MC) simulations. Significant discrepancies between MC and experimental values were found when lhe manufacturer parameters of lhe detector were used in lhe simulation. A complete Computerized Tomagraphy (CT) detector scan allowed to find the correct crystal dimensions and position inside the capsule. The computed efficiencies with the resulting detector model differed with the measured values no more than 10% in most of the energy range.