Evaluation of the detection and quantification limits in electroanalysis using two popular methods: application in the case study of paraquat determination


Autoria(s): da Silva, Osmair Benedito; Machado, Sergio Antonio Spinola
Contribuinte(s)

UNIVERSIDADE DE SÃO PAULO

Data(s)

06/11/2013

06/11/2013

2012

Resumo

In this work, the reduction reaction of paraquat herbicide was used to obtain analytical signals using electrochemical techniques of differential pulse voltammetry, square wave voltammetry and multiple square wave voltammetry. Analytes were prepared with laboratory purified water and natural water samples (from Mogi-Guacu River, SP). The electrochemical techniques were applied to 1.0 mol L-1 Na2SO4 solutions, at pH 5.5, and containing different concentrations of paraquat, in the range of 1 to 10 mu mol L-1, using a gold ultramicroelectrode. 5 replicate experiments were conducted and in each the mean value for peak currents obtained -0.70 V vs. Ag/AgCl yielded excellent linear relationships with pesticide concentrations. The slope values for the calibration plots (method sensitivity) were 4.06 x 10(-3), 1.07 x 10(-2) and 2.95 x 10(-2) A mol(-1) L for purified water by differential pulse voltammetry, square wave voltammetry and multiple square wave voltammetry, respectively. For river water samples, the slope values were 2.60 x 10(-3), 1.06 x 10(-2) and 3.35 x 10(-2) A mol(-1) L, respectively, showing a small interference from the natural matrix components in paraquat determinations. The detection limits for paraquat determinations were calculated by two distinct methodologies, i.e., as proposed by IUPAC and a statistical method. The values obtained with multiple square waves voltammetry were 0.002 and 0.12 mu mol L-1, respectively, for pure water electrolytes. The detection limit from IUPAC recommendations, when inserted in the calibration curve equation, an analytical signal (oxidation current) is smaller than the one experimentally observed for the blank solution under the same experimental conditions. This is inconsistent with the definition of detection limit, thus the IUPAC methodology requires further discussion. The same conclusion can be drawn by the analyses of detection limits obtained with the other techniques studied.

Secretaria de Educacao do Estado de Sao Paulo

Secretaria de Educacao do Estado de Sao Paulo

Conselho Nacional de Desenvolvimento Cientifico e Tecnologico (CNPq)

Conselho Nacional de Desenvolvimento Cientifico e Tecnologico (CNPq)

Identificador

ANALYTICAL METHODS, CAMBRIDGE, v. 4, n. 8, supl. 2, Part 3, pp. 2348-2354, DEC, 2012

1759-9660

http://www.producao.usp.br/handle/BDPI/41955

10.1039/c2ay25111f

http://dx.doi.org/10.1039/c2ay25111f

Idioma(s)

eng

Publicador

ROYAL SOC CHEMISTRY

CAMBRIDGE

Relação

ANALYTICAL METHODS

Direitos

openAccess

Copyright ROYAL SOC CHEMISTRY

Palavras-Chave #SQUARE-WAVE VOLTAMMETRY #CARBON-PASTE ELECTRODE #AQUEOUS-SOLUTION #NATURAL-WATER #HERBICIDE PARAQUAT #MICROELECTRODES #EUROPIUM(III) #CARBARYL #SAMPLES #RESIN #CHEMISTRY, ANALYTICAL #FOOD SCIENCE & TECHNOLOGY #SPECTROSCOPY
Tipo

article

original article

publishedVersion