Comparison of 3 modes of automated weaning from mechanical ventilation: A bench study
Contribuinte(s) |
UNIVERSIDADE DE SÃO PAULO |
---|---|
Data(s) |
07/11/2013
07/11/2013
2012
|
Resumo |
Purpose: Automated weaning modes are available in some mechanical ventilators, but no studies compared them hitherto. We compared the performance of 3 automated modes under standard and challenging situations. Methods: We used a lung simulator to compare 3 automated modes, adaptive support ventilation (ASV), mandatory rate ventilation (MRV), and Smartcare, in 6 situations, weaning success, weaning failure, weaning success with extreme anxiety, weaning success with Cheyne-Stokes, weaning success with irregular breathing, and weaning failure with ineffective efforts. Results: The 3 modes correctly recognized the situations of weaning success and failure, even when anxiety or irregular breathing were present but incorrectly recognized weaning success with Cheyne-Stokes. MRV incorrectly recognized weaning failure with ineffective efforts. Time to pressure support (PS) stabilization was shorter for ASV (1-2 minutes for all situations) and MRV (1-7 minutes) than for Smartcare (8-78 minutes). ASV had higher rates of PS oscillations per 5 minutes (4-15), compared with Smartcare (0-1) and MRV (0-12), except when extreme anxiety was present. Conclusions: Smartcare, ASV, and MRV were equally able to recognize weaning success and failure, despite the presence of anxiety or irregular breathing but performed incorrectly in the presence of Cheyne-Stokes. PS behavior over the time differs among modes, with ASV showing larger and more frequent PS oscillations over the time. Clinical studies are needed to confirm our results. (C) 2012 Elsevier Inc. All rights reserved. Fundacao de Amparo a Pesquisa do Estado de Sao Paulo (Fapesp), a government agency [2010/11545-0] Fundacao de Amparo a Pesquisa do Estado de Sao Paulo (Fapesp), a government agency |
Identificador |
JOURNAL OF CRITICAL CARE, PHILADELPHIA, v. 27, n. 6, supl. 4, Part 1-2, pp. 337-348, DEC, 2012 0883-9441 http://www.producao.usp.br/handle/BDPI/43230 10.1016/j.jcrc.2011.12.021 |
Idioma(s) |
eng |
Publicador |
W B SAUNDERS CO-ELSEVIER INC PHILADELPHIA |
Relação |
JOURNAL OF CRITICAL CARE |
Direitos |
closedAccess Copyright W B SAUNDERS CO-ELSEVIER INC |
Palavras-Chave | #VENTILATOR WEANING #MECHANICAL VENTILATORS #COMPUTER SIMULATION #VENTILATOR WEANING/METHODS #INTERMITTENT POSITIVE-PRESSURE VENTILATION #ADAPTIVE SUPPORT VENTILATION #RANDOMIZED CONTROLLED-TRIAL #BREATHING PATTERNS #CARDIAC-SURGERY #PRESSURE #LUNG #DURATION #CRITICAL CARE MEDICINE |
Tipo |
article original article publishedVersion |