Resumo

The use of an amperometric biosensor for the salicylate determination in blood serum is described. The biosensor is based on salicylate hydroxylase (EC 1.14.13.1) electropolymerized onto a glassy carbon-working electrode with polypyrrole and glutaraldehyde, to improve the biosensor lifetime. The hexacyanoferrate (II) was also incorporated to work as a redox mediator to minimize possible interferences. The salicylate is enzymatically converted to catechol, which is monitored amperometrically by its electrooxidation at +0.170 V versus SCE (saturated calomel electrode). Salicylate determination was carried out maintaining the ratio between β-NADH and salicylate at 4:1 (30°C). The amperometric response of the biosensor was linearly proportional to the salicylate concentration between 2.3 x 10-6 and 1.4 x 10-5 mol l- 1, in 0.1 mol l-1 phosphate buffer (pH 7.8), containing 0.1 mol l-1 KCl and 5.0 x 10-4 mol l-1 Na2H2EDTA, as supporting electrolyte. The recovery studies, in the presence of several interfering compounds, showed recoveries between 96.4 and 104.8%. The useful lifetime of the biosensor in the concentration range evaluated was at least 40 days, in continuous use. Blood serum samples analyzed by this biosensor showed a good correlation compared to the spectrophotometric method (Trinder) used as reference, presenting relative deviations lower than 7.0%. (C) 2000 Elsevier Science B.V.

Formato

547-557

Identificador

http://dx.doi.org/10.1016/S0039-9140(99)00311-2

Talanta. Amsterdam: Elsevier B.V., v. 51, n. 3, p. 547-557, 2000.

0039-9140

http://hdl.handle.net/11449/130628

10.1016/S0039-9140(99)00311-2

WOS:000085712300014

2-s2.0-0342905042

Idioma(s)

eng

Publicador

Elsevier B.V.

Relação

Talanta

Direitos

closedAccess

Palavras-Chave #Polypyrrole #Salicylate determination #Salicylate hydroxylase #Serum samples #Buffer #Carbon #Catechol #Electrolyte #Ferrocyanide #Glutaraldehyde #Mercurous chloride #Oxygenase #Phosphate #Polypyrrole #Reduced nicotinamide adenine dinucleotide #Salicylic acid #Amperometry #Biosensor #Blood analysis #Chemical composition #Controlled study #Drug determination #Electrode #Intermethod comparison #Oxidation #pH #Polymerization #Serum #Spectrophotometry #Temperature
Tipo

info:eu-repo/semantics/article