Analysis of Steroids using Solid Phase Microextraction-Gas Chromatography-Mass Spectrometry-Mass Spectrometry (SPME-GC-MS-MS)


Autoria(s): Chopra, Shilpi; Lima Gomes, Paulo Clairmont Feitosa de; Dhandapani, Ramkumar; Snow, Nicholas H.
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

15/05/2015

15/05/2015

2014

Resumo

Direct immersion SPME-GC-MS-MS was used for the analysis of steroids in water at part-per-trillion(ppt) and lower concentrations. The method was validated and extended to real sample analysis. The method were linear from 0.01 to 5 ng/ml with precision less than 10% relative standard deviation for a steroid mixture at 1 ng/ml. Limit of quantitation and limit of detection was found to be 200- 1200 pg/L and 30-200 pg/L respectively and recoveries ranged from 88-103 %. To understand the extraction efficiency of the fiber, a depletion study was performed. The fiber/ sample partition coefficients for the steroids were determined to be 1.0 x 104 to 1.5 x 104 . The extraction was performed without derivatization or the use of an internal standard. SPMEGC-MS-MS effectively demonstrated ultra-trace level detection of steroids in water.

Formato

105-116

Identificador

http://www.scientiachromatographica.com/doi/10.4322/sc.2014.024

Scientia Chromatographica, v. 6, n. 2, p. 105-116, 2014.

1984-4433

http://hdl.handle.net/11449/123538

http://dx.doi.org/10.4322/sc.2014.024

0049857977635051

Idioma(s)

por

Relação

Scientia Chromatographica

Direitos

closedAccess

Palavras-Chave #Steroids using Solid Phase Microextraction #SPME #Gas Chromatography-Mass Spectrometry-Mass Spectrometry #GC-MS-MS #SPMEGC-MS-MS #water analysis
Tipo

info:eu-repo/semantics/article