This paper presents a novel approach to the computed assessment of a mammographic phantom device. The approach shown here is fully automated and is based on the automatic selection of the region of interest, in the use of the discrete wavelet transform (DWT) and morphological operators to assess the quality of the American College of Radiology (ACR) mammographic phantom images. The algorithms developed here have succesfully scored 30 images obtained with different combinations of voltage applied to the tube and exposure and could notice the differences in the radiographs due to the different level of exposure to radiation. © 2013 Springer-Verlag.

Formato

811-814

Identificador

http://dx.doi.org/10.1007/978-3-642-29305-4_213

IFMBE Proceedings, v. 39 IFMBE, p. 811-814.

1680-0737

http://hdl.handle.net/11449/75121

10.1007/978-3-642-29305-4_213

2-s2.0-84876016115

Idioma(s)

eng

Relação

IFMBE Proceedings

Direitos

closedAccess

Palavras-Chave #accreditation phantom #computed image evaluation #computed radiography #mammography #quality control #American college of radiologies #Automatic selection #Computed radiography #Fully automated #Image evaluation #Mammographic phantoms #Morphological operator #Region of interest #Accreditation #Biomedical engineering #Discrete wavelet transforms #Physics #Quality control #Radiology #Mammography