Electrogravimetric Analysis by Quartz-Crystal Microbalance on the Consumption of the Neurotransmitter Acetylcholine by Acetylcholinesterase
Contribuinte(s) |
Universidade Estadual Paulista (UNESP) |
---|---|
Data(s) |
27/05/2014
27/05/2014
01/12/2012
|
Resumo |
Electrogravimetric analysis was performed on the consumption of the neurotransmitter Acetylcholine (ACh) by Acetylcholinesterase (AChE) in situ and in real time. Michaelis-Menten assumption was achieved by using an enzyme micro-reactor in which the total enzyme was anchored in a quartz crystal microbalance chip (QCM-chip) with a strategically engineered self-assembled monolayer (SAM) of alkanethiols, which can prevent diffusion-controlled or spatially restricted kinetics. The real-time frequency changes indicated the rate of the products formation from enzymatic reaction. The QCM-chip was tested showing that it could demonstrate AChE inhibition by physostigmine. © 2013 Copyright Taylor and Francis Group, LLC. |
Formato |
258-265 |
Identificador |
http://dx.doi.org/10.1080/00032719.2012.713065 Analytical Letters, v. 46, n. 2, p. 258-265, 2012. 0003-2719 1532-236X http://hdl.handle.net/11449/73799 10.1080/00032719.2012.713065 WOS:000312984400004 2-s2.0-84871859504 |
Idioma(s) |
eng |
Relação |
Analytical Letters |
Direitos |
closedAccess |
Palavras-Chave | #Acetyl cholinesterase #Acetylcholine #Neurotransmitter #Physostigmine #Quartz-crystal microbalance |
Tipo |
info:eu-repo/semantics/article |