Sistema para realização de medidas de luminescência com excitação por raios X
Contribuinte(s) |
Universidade Estadual Paulista (UNESP) |
---|---|
Data(s) |
27/05/2014
27/05/2014
22/08/2011
|
Resumo |
In this work is presented a versatile system for X-ray excited optical luminescence (XEOL) measurements. The apparatus was assembled from a sample holder connected to an optical fiber responsibly for the acquisition of the scintillation signal. The spectrum is registered with a CCD coupled in a spectrograph provided with diffraction gratings. The system performance was analyzed by exciting GdAlO3:Eu3+ 3.0 at.% with X-rays from a diffractometer and measuring the emission spectra. The system can be used to obtain precise and reliable spectroscopic properties of samples with various conformations without the loss of the required safety when dealing with ionizing radiations. |
Formato |
1057-1062 |
Identificador |
http://dx.doi.org/10.1590/S0100-40422011000600024 Quimica Nova, v. 34, n. 6, p. 1057-1062, 2011. 0100-4042 1678-7064 http://hdl.handle.net/11449/72607 10.1590/S0100-40422011000600024 S0100-40422011000600024 2-s2.0-80051719623 2-s2.0-80051719623.pdf |
Idioma(s) |
por |
Relação |
Química Nova |
Direitos |
openAccess |
Palavras-Chave | #Scintillators #Spectroscopy #X-ray phosphors |
Tipo |
info:eu-repo/semantics/article |