A novel potentiometric naproxenate ion sensor immobilized in a graphite matrix for determination of naproxen in pharmaceuticals


Autoria(s): Santini, Alberto O.; Oliveira, Jose Eduardo de; Pezza, Helena Redigolo; Pezza, Leonardo
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

27/05/2014

27/05/2014

09/08/2006

Resumo

The characteristics, performance, and application of an electrode, namely, Pt|Hg|Hg 2(NAP) 2| Graphite, where NAP stands for naproxenate ion, are described. This electrode responds to NAP with sensivity of (58.1± 0.9) mV decade -1 over the range 5.0 × 10 -5 - 1.0 × 10 -2 mol L -1 at pH 6.0-9.0 and a detection limit of 3.9 × 10 -5 mol L -1. The electrode is easily constructed at a relatively low cost with fast response time (within 10-35 s) and can be used for a period of 6 months without any considerable divergence in potentials. The proposed sensor displayed good selectivity for naproxen in the presence of several substances, especially concerning carboxylate and inorganic anions. It was used for the direct assay of naproxen in commercial tablets by means of the standard additions method. The analytical results obtained by using this electrode are in good agreement with those given by the United States Pharmacopeia procedure. ©2006 Sociedade Brasileira de Química.

Formato

785-791

Identificador

http://dx.doi.org/10.1590/S0103-50532006000400022

Journal of the Brazilian Chemical Society, v. 17, n. 4, p. 785-791, 2006.

0103-5053

http://hdl.handle.net/11449/69031

10.1590/S0103-50532006000400022

S0103-50532006000400022

WOS:000239616800021

2-s2.0-33746754680

2-s2.0-33746754680.pdf

Idioma(s)

eng

Relação

Journal of the Brazilian Chemical Society

Direitos

openAccess

Palavras-Chave #Naproxenate-sensitive electrode #Pharmaceutical formulations #Potentiometry
Tipo

info:eu-repo/semantics/article