Impedance e-tongue instrument for rapid liquid assessment


Autoria(s): Cabral, Flavio P. A.; Bergamo, Bruno B.; Dantas, Cleber A. R.; Riul, A.; Giacometti, Jose A.
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

20/05/2014

20/05/2014

01/02/2009

Resumo

Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)

Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

We present a compact and easy to handle instrument developed to perform rapid analysis of liquids utilizing an electronic tongue system. Briefly, the e-tongue used here is based on impedance measurements of an array of sensing units fabricated with ultrathin films of different materials deposited onto gold interdigitated electrodes. The instrument has the capability of measuring up to eight different sensor sets, each comprising an array of eight sensing units, and can perform a series of measurements in less than half of an hour. Additionally, there is a user-friendly software interface for instrument control, allowing the statistical correlation of samples using principal component analysis.

Formato

3

Identificador

http://dx.doi.org/10.1063/1.3084210

Review of Scientific Instruments. Melville: Amer Inst Physics, v. 80, n. 2, p. 3, 2009.

0034-6748

http://hdl.handle.net/11449/42323

10.1063/1.3084210

WOS:000263803500050

WOS000263803500050.pdf

Idioma(s)

eng

Publicador

American Institute of Physics (AIP)

Relação

Review of Scientific Instruments

Direitos

closedAccess

Palavras-Chave #electric impedance measurement #electronic tongues #gold #principal component analysis
Tipo

info:eu-repo/semantics/article