Carbamide Peroxide Determination in Tooth Whitening Using a Reagentless HRP-Biosensor


Autoria(s): Vidal, Isabelle Goncalves; Tescarollo Dias, Iara Lucia; de Oliveira Neto, Graciliano; de Vasconcellos Lanza, Marcos Roberto; Taboada Sotomayor, Maria Del Pilar
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

20/05/2014

20/05/2014

01/01/2009

Resumo

Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)

Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

A rapid and effective method for carbamide peroxide (CP) quantification in pharmaceuticals is proposed. The reagentless biosensor was prepared by using lyophilized turnip extract as horseradish peroxidase (HRP) enzyme source. The biosensor presented the best performance. The measurements were carried out in 0.1mol L-1 Pipes buffer (pH 6.0). At-100mV vs. Ag|AgCl, a linear response range from 3.0 to 25.0mmol L-1 was obtained. The quantification and detection limits were 1.2 and 0.4mmol L-1, respectively, and the response time was 0.5s. The biosensor repeatability, storage, and lifetime were excellent, allowing a satisfactory CP quantification in real pharmaceutical samples, when compared with those obtained by the official method.

Formato

352-365

Identificador

http://dx.doi.org/10.1080/00032710802507927

Analytical Letters. Philadelphia: Taylor & Francis Inc, v. 42, n. 2, p. 352-365, 2009.

0003-2719

http://hdl.handle.net/11449/41385

10.1080/00032710802507927

WOS:000263553400013

Idioma(s)

eng

Publicador

Taylor & Francis Inc

Relação

Analytical Letters

Direitos

closedAccess

Palavras-Chave #Amperometric biosensor #brassica napus L #carbamide peroxide #HRP #lyophilization procedure
Tipo

info:eu-repo/semantics/article