Arsenic as internal standard to correct for interferences in the determination of antimony by hydride generation in situ trapping graphite furnace atomic absorption spectrometry


Autoria(s): Freschi, C. D.; Freschi, GPG; Neto, JAG; Nobrega, J. A.; Oliveira, P. V.
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

20/05/2014

20/05/2014

30/06/2005

Resumo

The feasibility of using internal standardization (IS) to correct for interferences in hydride generation with in situ trapping in graphite furnace was evaluated. Arsenic was chosen as internal standard for Sb determination and Ir was used as permanent modifier. Fluctuations in the main parameters that affect the analytical results were minimized by IS and an effective contribution was verified in the studies of liquid phase interferences. Cobalt and Ni2+ were selected to illustrate the potential use of IS on the correction of interference by transition metals. The application of IS allows the Sb determination in samples containing up to 20-fold higher concentration of the Co2+ and Ni2+ when compared to the procedure without IS. The relative standard deviation of measurements varied from 0.3% to 0.7% and from 1.1% to 3.2% with and without IS, respectively. Recoveries within 92% and 107% of spiked aqueous solution containing Sb(III) and Sb(V) were found. (c) 2005 Elsevier B.V. All rights reserved.

Formato

759-763

Identificador

http://dx.doi.org/10.1016/j.sab.2005.03.004

Spectrochimica Acta Part B-atomic Spectroscopy. Oxford: Pergamon-Elsevier B.V., v. 60, n. 5, p. 759-763, 2005.

0584-8547

http://hdl.handle.net/11449/39590

10.1016/j.sab.2005.03.004

WOS:000230022200029

Idioma(s)

eng

Publicador

Elsevier B.V.

Relação

Spectrochimica Acta Part B-atomic Spectroscopy

Direitos

closedAccess

Palavras-Chave #internal standardization #atomic absorption spectrometry #hydride generation #in situ trapping in graphite furnace #antimony #arsenic
Tipo

info:eu-repo/semantics/article