Dissolved organic carbon in rainwater: Glassware decontamination and sample preservation and volatile organic carbon
Contribuinte(s) |
Universidade Estadual Paulista (UNESP) |
---|---|
Data(s) |
20/05/2014
20/05/2014
01/12/2007
|
Resumo |
The efficiency of different methods for the decontamination of glassware used for the analysis of dissolved organic carbon (DOC) was tested using reported procedures as well as new ones proposed in this work. A Fenton solution bath (1.0 mmol L-1 Fe2+ and 100 mmol L-1 H2O2) for 1 h or for 30 min employing UV irradiation showed to combine simplicity, low cost and high efficiency. Using the optimized cleaning procedure, the DOC for stored UV-irradiated ultrapure water reached concentrations below the limit of detection (0.19 mu mol C L-1). Filtered (0.7 mu m) rain samples maintained the DOC integrity for at least 7 days when stored at 4 degrees C. The volatile organic carbon (VOC) fraction in the rain samples collected at two sites in São Paulo state (Brazil) ranged from 0% to 56% of their total DOC content. Although these high-VOC concentrations may be derived from the large use of ethanol fuel in Brazil, our results showed that when using the high-temperature catalytic oxidation technique, it is essential to measure DOC rather than non-purgeble organic carbon to estimate organic carbon, since rainwater composition can be quite variable, both geographically and temporally. (C) 2007 Elsevier Ltd. All rights reserved. |
Formato |
8924-8931 |
Identificador |
http://dx.doi.org/10.1016/j.atmosenv.2007.08.017 Atmospheric Environment. Oxford: Pergamon-Elsevier B.V., v. 41, n. 39, p. 8924-8931, 2007. 1352-2310 http://hdl.handle.net/11449/33873 10.1016/j.atmosenv.2007.08.017 WOS:000252260400008 |
Idioma(s) |
eng |
Publicador |
Elsevier B.V. |
Relação |
Atmospheric Environment |
Direitos |
closedAccess |
Palavras-Chave | #glassware decontamination for DOC #rainwater DOC storage #wet deposition #carbon flux #non-purgeble organic carbon |
Tipo |
info:eu-repo/semantics/article |