Cathodic stripping voltammetric determination of cefaclor in pharmaceutical formulations


Autoria(s): Rodrigues, LNC; Fogg, A. G.; Zanoni, Maria Valnice Boldrin
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

20/05/2014

20/05/2014

01/01/1999

Resumo

A sensitive method is described for the determination of cefaclor by cathodic stripping voltammetry at the hanging mercury drop electrode. cefaclor is accumulated at the electrode surface as a mercury salt, which is reduced at -0.67 V. The optimum accumulation potential and accumulation time were +0.15 V and up to 180 s, respectively. Linear calibration graphs were obtained between 3.9 mu g.L-1 to 39 mu g.L-1 and the limit of determination was evaluated to be 1.9 mu g.L-1. The method was applied successfully to the determination of cefaclor in pharmaceutical formulations.

Formato

97-109

Identificador

http://dx.doi.org/10.1080/00032719908542601

Analytical Letters. New York: Marcel Dekker Inc., v. 32, n. 1, p. 97-109, 1999.

0003-2719

http://hdl.handle.net/11449/32855

10.1080/00032719908542601

WOS:000078552100008