A new potentiometric ibuprofenate ion sensor immobilized in a graphite matrix for determination of ibuprofen in tablets


Autoria(s): Santini, A. O.; Oliveira, Jose Eduardo de; Pezza, Helena Redigolo; Pezza, Leonardo
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

20/05/2014

20/05/2014

01/09/2006

Resumo

The characteristics, performance, and application of an electrode, namely, Pt vertical bar Hg vertical bar Hg-2(IBP)(2)vertical bar Graphite, where IBP stands for ibuprofenate ion, are described. This electrode responds to IBP with sensitivity of (58.6 +/- 0.9) mV decade 1 over the range 5.0 x 10(-5)-1.0 x 10(-1) mol L-1 at pH 6.0-9.0 and a detection limit of 3.8 x 10(-5) mol L-1. The electrode is easily constructed at a relatively low cost with fast response time (within 1530 s) and can be used for a period of 5 months without any considerable divergence in potentials. The proposed sensor displayed good selectivity for ibuprofen in the presence of several substances, especially concerning carboxylate and inorganic anions. It was used for the direct assay of ibuprofen in commercial tablets by means of the standard additions method. The analytical results obtained by using this electrode are in good agreement with those given by the United States Pharmacopeia procedure. (c) 2006 Elsevier B.V. All rights reserved.

Formato

44-49

Identificador

http://dx.doi.org/10.1016/j.microc.2006.04.007

Microchemical Journal. Amsterdam: Elsevier B.V., v. 84, n. 1-2, p. 44-49, 2006.

0026-265X

http://hdl.handle.net/11449/32270

10.1016/j.microc.2006.04.007

WOS:000240428800008

Idioma(s)

eng

Publicador

Elsevier B.V.

Relação

Microchemical Journal

Direitos

closedAccess

Palavras-Chave #ibuprofenate-sensitive electrode #potentiometry #tablet formulations
Tipo

info:eu-repo/semantics/article