Construção de uma célula-autoclave para medidas eletroquímicas a altas temperaturas


Autoria(s): Ferreira, Elivelton Alves; Polachini, Ferminio Cesar; Fugivara, Cecilio Sadao; Benedetti, Assis Vicente
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

20/05/2014

20/05/2014

01/09/2011

Resumo

Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)

Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

Measurements at high temperature using liquid solutions require special cells and materials which are able to support the temperature and pressure developed inside. The constructed cell was designed to support pressures up to 20 bar, temperatures relatively high up to around 200 ºC, depending on the pressure developed inside the system. It also supports aggressive solutions since its inner wall is made of Teflon. The electrolyte has no contact with the metallic body of the cell. Then, it is supposed that this work represents a great contribution to the electrochemical studies of materials in solutions at high pressure and temperature

Formato

1647-1650

Identificador

http://dx.doi.org/10.1590/S0100-40422011000900027

Química Nova. Sociedade Brasileira de Química, v. 34, n. 9, p. 1647-1650, 2011.

0100-4042

http://hdl.handle.net/11449/25474

10.1590/S0100-40422011000900027

S0100-40422011000900027

WOS:000297313200027

S0100-40422011000900027.pdf

Idioma(s)

por

Publicador

Sociedade Brasileira de Química

Relação

Química Nova

Direitos

openAccess

Palavras-Chave #electrochemical cell-autoclave #electrochemical polarization #high pressure electrochemical cell
Tipo

info:eu-repo/semantics/article