Determination of Cd(II), Cu(II) and Ni(II) in aqueous samples by ICP-OES after on-line preconcentration in column packed with silica modified with 2-aminothiazole
Contribuinte(s) |
Universidade Estadual Paulista (UNESP) |
---|---|
Data(s) |
20/05/2014
20/05/2014
01/01/2008
|
Resumo |
This work describes the synthesis and characterization of 2-aminothiazole modified silica gel (SiAT) and the studies of adsorption and pre-concentration (in batch and using a flow-injection system coupled with optical emission spectrometer) of Cd(II), Cu(II) and Ni(II) in aqueous medium. The adsorption capacity for each metal ions in mmolg(-1) was: Cu(II) = 1.18, Ni(II) = 1.15 and Cd(II) = 1.10. The results obtained in the flow experiments showed about 100% of recovering of the metal ions adsorbed in a mini-column packed with 100 mg of SiAT, using 100 mu L of 2.0 mol L-1 HCl solution as eluent. The quantitative sorption-desorption of the metal ions made possible the application of a flow-injection system in the pre-concentration and quantification by ICP-OES of metal ions at trace level in natural water samples. |
Formato |
203-209 |
Identificador |
http://dx.doi.org/10.1007/s00604-007-0845-5 Microchimica Acta. Wien: Springer Wien, v. 160, n. 1-2, p. 203-209, 2008. 0026-3672 http://hdl.handle.net/11449/17643 10.1007/s00604-007-0845-5 WOS:000252484600020 |
Idioma(s) |
eng |
Publicador |
Springer Wien |
Relação |
Microchimica Acta |
Direitos |
closedAccess |
Palavras-Chave | #pre-concentration #2-aminothiazole #flow-injection system #sol-gel method #ICP-OES |
Tipo |
info:eu-repo/semantics/article |