Controle de qualidade e dosimetria em equipamentos de tomografia computadorizada


Autoria(s): Pina, Diana Rodrigues de; Duarte, Sérgio Barbosa; Netto, Thomaz Ghilardi; Morceli, José; Carbi, Eros Duarte Ortigoso; Souza, Rafael Toledo Fernandes de; Costa Neto, André; Ribeiro, Sérgio Marrone
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

20/05/2014

20/05/2014

01/06/2009

Resumo

OBJETIVO: Avaliação de condições dos equipamentos e dosimetria em setores de tomografia computadorizada utilizando protocolos de cabeça, abdome e coluna lombar em pacientes adultos (em três equipamentos distintos) e pediátricos com até um ano e meio de vida (em um dos equipamentos avaliados). MATERIAIS E MÉTODOS: Foram estimados o índice de dose em tomografia computadorizada e a dose média em cortes múltiplos, em exames com pacientes adultos, em três distintos equipamentos. Ainda foram estimadas as doses na superfície de entrada e as doses absorvidas em exame de cabeça para pacientes adultos e pediátricos em um dos equipamentos avaliados. RESULTADOS: Foram realizados testes de controle de qualidade, mecânicos, demonstrando que os equipamentos satisfazem as especificações de uso estabelecidas pelas normas vigentes. Os resultados da dosimetria mostraram que valores de dose média em cortes múltiplos excederam em até 109,0% os valores de níveis de referência, apresentando consideráveis variações entre os equipamentos avaliados neste estudo. As doses absorvidas obtidas com protocolos pediátricos são inferiores aos de pacientes adultos, apresentando redução de até 51,0% na tireoide. CONCLUSÃO: Neste estudo foram avaliadas as condições de operação de três equipamentos tomográficos, estabelecendo quais parâmetros devem ser trabalhados para a implantação de um programa de controle de qualidade nas instituições onde esta pesquisa foi desenvolvida.

OBJECTIVE: Evaluation of equipment conditions and dosimetry in computed tomography services utilizing protocols for head, abdomen, and lumbar spine in adult patients (in three different units) and pediatric patients up to 18 months of age (in one of the units evaluated). MATERIALS and METHODS: Computed tomography dose index and multiple-scan average dose were estimated in studies of adult patients with three different units. Additionally, entrance surface doses as well as absorbed dose were estimated in head studies for both adult and pediatric patients in a single computed tomography unit. RESULTS: Mechanical quality control tests were performed, demonstrating that computed tomography units comply with the equipment-use specifications established by the current standards. Dosimetry results have demonstrated that the multiple-scan average dose values were in excess of up to 109.0% the reference levels, presenting considerable variation amongst the computed tomography units evaluated in the present study. Absorbed doses obtained with pediatric protocols are lower than those with adult patients, presenting a reduction of up to 51.0% in the thyroid gland. CONCLUSION: The present study has analyzed the operational conditions of three computed tomography units, establishing which parameters should be set for the deployment of a quality control program in the institutions where this study was developed.

Formato

171-177

Identificador

http://dx.doi.org/10.1590/S0100-39842009000300010

Radiologia Brasileira. Colégio Brasileiro de Radiologia e Diagnóstico por Imagem, v. 42, n. 3, p. 171-177, 2009.

0100-3984

http://hdl.handle.net/11449/11499

10.1590/S0100-39842009000300010

S0100-39842009000300010

S0100-39842009000300010.pdf

Idioma(s)

por

Publicador

Colégio Brasileiro de Radiologia e Diagnóstico por Imagem

Relação

Radiologia Brasileira

Direitos

openAccess

Palavras-Chave #Equipamento tomográfico #Controle de qualidade #Dosimetria #Redução de dose #Computed tomography equipment #Quality control #Dosimetry #Dose reduction
Tipo

info:eu-repo/semantics/article