Application of solid sampling device for direct determination of chromium and nickel in lubricating oils by graphite furnace atomic absorption spectrometry
Contribuinte(s) |
UNIVERSIDADE DE SÃO PAULO |
---|---|
Data(s) |
20/10/2012
20/10/2012
2008
|
Resumo |
One method using a solid sampling device for the direct determination of Cr and Ni in fresh and used lubricating oils by graphite furnace atomic absorption spectrometry are proposed. The high organic content in the samples was minimized using a digestion step at 400 degrees C in combination with an oxidant mixture 1.0% (v v(-1)) HNO3+15% (v v(-1)) H2O2+0.1% (m v(-1)) Triton X-100 for the in situ digestion. The 3-field mode Zeeman-effect allowed the spectrometer calibration up to 5 ng of Cr and Ni. The quantification limits were 0.86 mu g g(-1) for Cr and 0.82 mg g(-1) for Ni, respectively. The analysis of reference materials showed no statistically significant difference between the recommended values and those obtained by the proposed methods. |
Identificador |
ANALYTICAL LETTERS, v.41, n.9, p.1547-1554, 2008 0003-2719 http://producao.usp.br/handle/BDPI/31359 10.1080/00032710802122099 |
Idioma(s) |
eng |
Publicador |
TAYLOR & FRANCIS INC |
Relação |
Analytical Letters |
Direitos |
restrictedAccess Copyright TAYLOR & FRANCIS INC |
Palavras-Chave | #atomic absorption #chromium #direct oil sampling #nickel #WEAR METALS #CU #Chemistry, Analytical |
Tipo |
article proceedings paper publishedVersion |