Electrochemical Behavior of biosensor by determination the catechol and phenolic compounds using the enzyme PPO
Contribuinte(s) |
UNIVERSIDADE DE SÃO PAULO |
---|---|
Data(s) |
20/10/2012
20/10/2012
2009
|
Resumo |
An electrochemical biosensor using poly-phenol oxidasa (PPO) was constructed for the determination of phenolic compounds. The PPO employed with enzyme, it was obtained from Archontophoenix Cunninghamiana. The biosensor showed range of linearity in the range of 1 x 10(-3) to 1 x 10(-4) mol/L and a detection limit of 1 x 10(-4) mol/L. The optimal pH was 6,7 in medium phosphate buffer. The lifetime of the biosensors was 1 months, stored in phosphate buffer solution 0.1 mol/L to ambient temperature. |
Identificador |
AFINIDAD, v.66, n.544, p.488-492, 2009 0001-9704 |
Idioma(s) |
por |
Publicador |
ASOC QUIMICOS |
Relação |
Afinidad |
Direitos |
closedAccess Copyright ASOC QUIMICOS |
Palavras-Chave | #Carbon paste biosensor #catechol #polyphenol oxidase #GLASSY-CARBON ELECTRODE #AMPEROMETRIC BIOSENSORS #ASCORBIC-ACID #EPINEPHRINE #DOPAMINE #FILMS #Chemistry, Multidisciplinary |
Tipo |