A portable Compton spectrometer for clinical X-ray beams in the energy range 20-150 keV


Autoria(s): VIEIRA, A. A.; Linke, Arlene; Yoshimura, Elisabeth Mateus; TERINI, R. A.; Herdade, Silvio Bruni
Contribuinte(s)

UNIVERSIDADE DE SÃO PAULO

Data(s)

20/10/2012

20/10/2012

2011

Resumo

Primary beam spectra were obtained for an X-ray industrial equipment (40-150 kV), and for a clinical mammography apparatus (25-35 kV) from beams scattered at angles close to 90 degrees, measured with a CdTe Compton spectrometer. Actual scattering angles were determined from the Compton energy shift of characteristic X-rays or spectra end-point energy. Evaluated contribution of coherent scattering amounts to more than 15% of fluence in mammographic beams. This technique can be used in clinical environments. (C) 2010 Elsevier Ltd. All rights reserved.

Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)

FAPESP

Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

CNPq

Identificador

APPLIED RADIATION AND ISOTOPES, v.69, n.2, p.350-357, 2011

0969-8043

http://producao.usp.br/handle/BDPI/29100

10.1016/j.apradiso.2010.10.011

http://dx.doi.org/10.1016/j.apradiso.2010.10.011

Idioma(s)

eng

Publicador

PERGAMON-ELSEVIER SCIENCE LTD

Relação

Applied Radiation and Isotopes

Direitos

restrictedAccess

Copyright PERGAMON-ELSEVIER SCIENCE LTD

Palavras-Chave #X-ray spectrometry #Compton spectrometry #Klein-Nishina cross-section #CdTe detector #Radiology #MONTE-CARLO-SIMULATION #SPECTRA #PHOTODIODE #DETECTOR #Chemistry, Inorganic & Nuclear #Nuclear Science & Technology #Radiology, Nuclear Medicine & Medical Imaging
Tipo

article

original article

publishedVersion