Spatial calibration of an optical see-through head-mounted display
Data(s) |
2008
|
---|---|
Resumo |
We present here a method for calibrating an optical see-through Head Mounted Display (HMD) using techniques usually applied to camera calibration (photogrammetry). Using a camera placed inside the HMD to take pictures simultaneously of a tracked object and features in the HMD display, we could exploit established camera calibration techniques to recover both the intrinsic and extrinsic properties of the~HMD (width, height, focal length, optic centre and principal ray of the display). Our method gives low re-projection errors and, unlike existing methods, involves no time-consuming and error-prone human measurements, nor any prior estimates about the HMD geometry. |
Formato |
text |
Identificador |
http://centaur.reading.ac.uk/2062/1/gfg2008prepint.pdf Gilson, S. J., Fitzgibbon, A. W. and Glennerster, A. <http://centaur.reading.ac.uk/view/creators/90000482.html> (2008) Spatial calibration of an optical see-through head-mounted display. Journal of Neuroscience Methods, 173 (1). pp. 140-146. ISSN 0165-0270 doi: 10.1016/j.jneumeth.2008.05.015 <http://dx.doi.org/10.1016/j.jneumeth.2008.05.015> |
Idioma(s) |
en |
Publicador |
Elsevier |
Relação |
http://centaur.reading.ac.uk/2062/ http://dx.doi.org/10.1016/j.jneumeth.2008.05.015 10.1016/j.jneumeth.2008.05.015 10.1016/j.jneumeth.2008.05.015 |
Palavras-Chave | #152 Perception, movement, emotions, drives |
Tipo |
Article PeerReviewed |
Direitos |