Non-invasive Studies on Age Related Parameters Using a Blood Volume Pulse Sensor


Autoria(s): Radhakrishnan, P; Jayasree, V K; Sandhya, T V
Data(s)

13/07/2010

13/07/2010

2008

Resumo

A non-invasive technique is implemented to measure a parameter which is closely related to the distensibility of large arteries, using the second derivative of the infrared photoplethysmographic waveform. Thirty subjects within the age group of 20-61 years were involved in this pilot study. Two new parameters, namely the area of the photoplethysmographic waveform under the systolic peak, and the ratio of the time delay between the systolic and the diastolic peaks and the time period of the waveform ( T/T) were studied as a function of age. It was found that while the parameter which is supposed to be a marker of distensibility of large arteries and T /T values correlate negatively with age, the area under the systolic peak correlates positively with age. The results suggest that the derived parameters could provide a simple, non-invasive means for studying the changes in the elastic properties of the vascular system as a function of age.

Identificador

International School of Photonics and CELOS, Cochin University of Science and Technology

measuremenr science review, Volume 8, Section 2, No. 4, 2008

http://dyuthi.cusat.ac.in/purl/1794

Idioma(s)

en

Publicador

Measurement Science Review

Palavras-Chave #vascular compliance #photoplethysmograph #systolic peak #second derivative
Tipo

Working Paper