Optimization and validation of a methodology to determine total arsenic, As(III) and As(V), in water samples, through graphite furnace atomic absorption spectrometry


Autoria(s): Santos,Lisia Maria Gobbo; Jacob,Silvana do Couto
Data(s)

01/03/2009

Resumo

The Graphite furnace atomic absorption spectrometry (GF AAS) was the technique chosen by the inorganic contamination laboratory (INCQ/ FIOCRUZ) to be validated and applied in routine analysis for arsenic detection and quantification. The selectivity, linearity, sensibility, detection, and quantification limits besides accuracy and precision parameters were studied and optimized under Stabilized Temperature Platform Furnace (STPF) conditions. The limit of detection obtained was 0.13 µg.L-1 and the limit of quantification was 1.04 µg.L-1, with an average precision, for total arsenic, less than 15% and an accuracy of 96%. To quantify the chemical species As(III) and As(V), an ion-exchange resin (Dowex 1X8, Cl- form) was used and the physical-chemical parameters were optimized resulting in a recuperation of 98% of As(III) and of 90% of As(V). The method was applied to groundwater, mineral water, and hemodialysis purified water samples. All results obtained were lower than the maximum limit values established by the legal Brazilian regulations, in effect, 50, 10, and 5 µg.L-1 para As total, As(III) e As(V), respectively. All results were statistically evaluated.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0101-20612009000100019

Idioma(s)

en

Publicador

Sociedade Brasileira de Ciência e Tecnologia de Alimentos

Fonte

Food Science and Technology (Campinas) v.29 n.1 2009

Palavras-Chave #arsenic #water #graphite furnace atomic absorption spectrometry #speciation #validation
Tipo