Potentiometric determination of Minoxidil in topical use pharmaceutical samples
Data(s) |
01/01/2009
|
---|---|
Resumo |
A potentiometric titration method for the determination of minoxidil based on its redox reaction with K2Cr2O7 is described. The best results were observed using 1.00 x 10-3 mol L-1 K2Cr2O7 and 1.00 x 10-2 mol L-1 minoxidil solutions, and the minoxidil as titrant in 2.00 mol L-1 H2SO4 medium. The method was applied to commercial samples and compared with the results from a chromatographic procedure. Recoveries from 97.4 to 98.7 % were observed depending on the sample. Comparison with the chromatographic procedure reveled agreement within 90% confidence level. |
Formato |
text/html |
Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-46702009000200003 |
Idioma(s) |
en |
Publicador |
Fundação Editora da Universidade Estadual Paulista Júlio de Mesquita Filho - UNESP |
Fonte |
Eclética Química v.34 n.2 2009 |
Palavras-Chave | #Minoxidil determination #titration #potenciometry |
Tipo |