Potentiometric determination of Minoxidil in topical use pharmaceutical samples


Autoria(s): Sousa,Rogério Adelino de; Cavalheiro,Éder Tadeu Gomes
Data(s)

01/01/2009

Resumo

A potentiometric titration method for the determination of minoxidil based on its redox reaction with K2Cr2O7 is described. The best results were observed using 1.00 x 10-3 mol L-1 K2Cr2O7 and 1.00 x 10-2 mol L-1 minoxidil solutions, and the minoxidil as titrant in 2.00 mol L-1 H2SO4 medium. The method was applied to commercial samples and compared with the results from a chromatographic procedure. Recoveries from 97.4 to 98.7 % were observed depending on the sample. Comparison with the chromatographic procedure reveled agreement within 90% confidence level.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-46702009000200003

Idioma(s)

en

Publicador

Fundação Editora da Universidade Estadual Paulista Júlio de Mesquita Filho - UNESP

Fonte

Eclética Química v.34 n.2 2009

Palavras-Chave #Minoxidil determination #titration #potenciometry
Tipo

journal article