Determination of atenolol in pharmaceutical formulations by diffuse reflectance spectroscopy


Autoria(s): Gotardo,M. A.; Sequinel,R.; Pezza,L.; Pezza,H. R.
Data(s)

01/12/2008

Resumo

A simple analytical method for quantification of atenolol in pharmaceutical formulations by diffuse reflectance spectroscopy is described. The method is based on the reaction, on the filter paper surface, between the drug and p-chloranil producing a colored compound. The best reaction conditions were obtained with 20 µL of atenolol solution and 20 µL of p-chloranil. All reflectance measurements were carried out at 550 nm and the linear range was from 1.13x10-2 to 7.88x10-2 mol L-1 (r = 0.9992). The limit of detection was 2.80 x 10-3 mol L-1. The proposed method was successfully applied to analysis of different commercial brands of pharmaceutical formulations and the results obtained by the proposed method were in good agreement with those obtained using the British Pharmacopoeia method.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-46702008000400001

Idioma(s)

en

Publicador

Fundação Editora da Universidade Estadual Paulista Júlio de Mesquita Filho - UNESP

Fonte

Eclética Química v.33 n.4 2008

Palavras-Chave #diffuse reflectance spectroscopy #atenolol #p-chloranil #pharmaceutical formulations
Tipo

journal article