Determinação de nimesulida por análise por injeção em fluxo com detecção amperométrica de múltiplos pulsos
Data(s) |
01/01/2013
|
---|---|
Resumo |
A simple and fast method for the determination of nimesulide (NI) using flow injection analysis with multiple-pulse amperometric (FIA-MPA) detection at a boron-doped diamond (BDD) electrode was developed. The method was based mainly on the application of a four-potential waveform, E1(det) = -0.8 V / 30 ms, E2(det) = 0.6 V / 30 ms, E3(det) = -0.4 V / 30 ms and E4(cleaning) = -0.45 V / 300 ms versus Ag/AgCl (3.0 mol L-1 KCl). NI was detected at three different electrode potentials, at which the nitro group undergoes different redox reactions. The proposed method was selective and sensitive (detection limit of 81.0 nmol L-1), and successfully applied for the determination of NI in pharmaceutical formulations, yielding similar results to those obtained by the reference method. |
Formato |
text/html |
Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422013000900004 |
Idioma(s) |
pt |
Publicador |
Sociedade Brasileira de Química |
Fonte |
Química Nova v.36 n.9 2013 |
Palavras-Chave | #pulse amperometric detection #nimesulide #boron-doped diamond |
Tipo |
journal article |