Determination of levodopa in human plasma by high performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS): application to a bioequivalence study


Autoria(s): Martins,Heliana F.; Pinto,Douglas P.; Nascimento,Viviane de A.; Marques,Marlice A. S.; Amendoeira,Fábio C.
Data(s)

01/01/2013

Resumo

A sensitive, accurate and simple method using HPLC-MS/MS was developed and validated for levodopa quantitation in human plasma. Analysis was achieved on a pursuit® C18 analytical column (5 µm; 150 x 4.6 mm i.d.) using a mobile phase (methanol and water , 90:10, v/v) containing formic acid 0.5% v/v, after extracting the samples using a simple protein plasma precipitation with perchloric acid. The developed method was validated in accordance with ANVISA guidelines and was successfully applied to a bioequivalence study in 60 healthy volunteers demonstrating the feasibility and reliability of the proposed method.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422013000100028

Idioma(s)

en

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.36 n.1 2013

Palavras-Chave #levodopa #HPLC-ESI-MS/MS #bioequivalence
Tipo

journal article