Multi-component quantitation of loratadine, pseudoephedrine and paracetamol in plasma and pharmaceutical formulations with liquid chromatography-tandem mass spectrometry utilizing a monolithic column


Autoria(s): Abro,Kamran; Memon,Najma; Bhanger,M. I.; Perveen,Shahnaz; Kandhro,Aftab
Data(s)

01/01/2012

Resumo

The purpose of this study was to develop a rapid, simple and sensitive quantitation method for pseudoephedrine (PSE), paracetamol (PAR) and loratadine (LOR) in plasma and pharmaceuticals using liquid chromatography-tandem mass spectrometry with a monolithic column. Separation was achieved using a gradient composition of methanol-0.1% formic acid at a flow rate of 1.0 mL min-1. Mass spectral transitions were recorded in SRM mode. System validation was evaluated for precision, specificity and linearity. Limit of detection for pseudoephedrine, paracetamol, and loratadine were determined to be 3.14, 1.86 and 1.44 ng mL-1, respectively, allowing easy determination in plasma with % recovery of 93.12 to 101.56%.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012001000012

Idioma(s)

en

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.35 n.10 2012

Palavras-Chave #antihistamines #decongestants #liquid chromatography-tandem MS
Tipo

journal article