Determinação simultânea de ácido ascórbico e ácido acetilsalicílico usando análise por injeção em fluxo com detecção amperométrica pulsada


Autoria(s): Miranda,Joyce A. T. de; Cunha,Rafael R.; Gimenes,Denise T.; Munoz,Rodrigo A. A.; Richter,Eduardo M.
Data(s)

01/01/2012

Resumo

A simple flow system with multiple pulse amperometric detection using a single working electrode is proposed for simultaneous determination of ascorbic (AA) and acetylsalicylic (AAS) acids in pharmaceutical formulations. The procedure is based on application of two potential pulses: 0.90 V/50 ms: oxidation and determination of AA without the interference of AAS; 1.35 V/50 ms: oxidation of both compounds and quantification of AAS by current subtraction using a correction factor. Sampling rate was estimated as 125 injections per hour and the limits of detection were 0.17 and 0.16 µmol L-1 for AA and AAS, respectively. Results for commercial samples agreed with those obtained using HPLC.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000700029

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.35 n.7 2012

Palavras-Chave #simultaneous determination #FIA #multiple pulse amperometry
Tipo

journal article