Determinação rápida de fármacos básicos em plasma por cromatografia a gás com detector de nitrogênio e fósforo


Autoria(s): Lizot,Lilian Feltraco; Silva,Luana Letícia da; Spaniol,Bárbara; Antunes,Marina Venzon; Linden,Rafael; Lanaro,Rafael
Data(s)

01/01/2012

Resumo

A simple and fast method for determination of 40 basic drugs in human plasma employing gas-chromatography with nitrogen-phosphorus detection was developed and validated. Drugs were extracted from 800 µL of plasma with 250 µL of butyl acetate at basic pH. Aliquots of the organic extract were directly injected on a column with methylsilicone stationary phase. Total chromatographic run time was 25 min. All compounds were detected in concentrations ranging from therapeutic to toxic levels, with intermediate precision CV% below 11.2 and accuracy in the range of 92-114%.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000600029

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.35 n.6 2012

Palavras-Chave #GC-NPD #emergency toxicology #drug screening
Tipo

journal article