Simultaneous determination of non-steroidal anti-inflammatory drugs in pharmaceutical formulations and human serum by reversed phase high performance liquid chromatography


Autoria(s): Nawaz,Muhammad
Data(s)

01/01/2012

Resumo

A rapid and sensitive method using high performance liquid chromatography has been developed and validated for the simultaneous determination of non-steroidal anti-inflammatory drugs (NSAIDs) in pharmaceutical formulations and human serum. Six NSAIDs including: naproxen sodium, diclofenac sodium, meloxicam, flurbiprofen, tiaprofenic and mefenamic acid were analyzed simultaneously in presence of ibuprofen as internal standard on Mediterranea C18 (5 µm, 250 x 0.46 mm) column. Mobile phase comprised of methanol: acetonitrile: H2O (60:20:20, v/v; pH 3.35) and pumped at a flow rate of 1 mL min-1 using 265 nm UV detection. The method was linear over a concentration range of 0.25-50 µg mL-1 (r² = 0.9999).

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000500014

Idioma(s)

en

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.35 n.5 2012

Palavras-Chave #NSAIDs #serum #reversed phase high performance liquid chromatography
Tipo

journal article