Desenvolvimento e validação de método analítico para quantificação do fármaco bevacizumabe por cromatografia a líquido de alta eficiência
| Data(s) |
01/01/2012
|
|---|---|
| Resumo |
In this study, an analytical method was developed and validated for quantitation of the drug bevacizumab (Avastin®) by high performance liquid chromatography (HPLC). The HPLC column was a BioSuite 250® HR SEC, 300 x 7.8 mm x 5 µm (Waters, USA). The mobile phase consisted of phosphate buffered saline (PBS). The results revealed that the method was specific, precise, accurate, robust and linear (r² = 0.998) from 5 to 75 µg mL-1. Therefore, this method can be used in drug release studies or in quality control ampoules of the drug. |
| Formato |
text/html |
| Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000300029 |
| Idioma(s) |
pt |
| Publicador |
Sociedade Brasileira de Química |
| Fonte |
Química Nova v.35 n.3 2012 |
| Palavras-Chave | #bevacizumab #high performance liquid chromatography #validation |
| Tipo |
journal article |