Performance characteristics of high performance liquid chromatography, first order derivative UV spectrophotometry and bioassay for fluconazole determination in capsules


Autoria(s): Corrêa,Josilene Chaves Ruela; Reichman,Camila; Salgado,Hérida Regina Nunes; Vianna-Soares,Cristina Duarte
Data(s)

01/01/2012

Resumo

The bioassay, first order derivative UV spectrophotometry and chromatographic methods for assaying fluconazole capsules were compared. They have shown great advantages over the earlier published methods. Using the first order derivative, the UV spectrophotometry method does not suffer interference of excipients. Validation parameters such as linearity, precision, accuracy, limit of detection and limit of quantitation were determined. All methods were linear and reliable within acceptable limits for antibiotic pharmaceutical preparations being accurate, precise and reproducible. The application of each method as a routine analysis should be investigated considering cost, simplicity, equipment, solvents, speed, and application to large or small workloads.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000300017

Idioma(s)

en

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.35 n.3 2012

Palavras-Chave #fluconazole #liquid chromatography #bioassay
Tipo

journal article