Derivative spectrophotometric method for determination of acyclovir in polymeric nanoparticles
| Data(s) |
01/01/2012
|
|---|---|
| Resumo |
A derivative spectrophotometric method was validated for quantification of acyclovir in poly (n-butylcyanoacrylate) (PBCA) nanoparticles. Specificity, linearity, precision, accuracy, recovery, detection (LOD) and quantification (LOQ) limits were established for method validation. First-derivative at 295.2 nm eliminated interferences from nanoparticle ingredients and presented linearity for acyclovir concentrations ranging from 1.25 to 40.0 µg/mL (r = 0.9999). Precision and accuracy data demonstrated good reproducibility. Recovery ranged from 99.3 to 101.2. LOD was 0.08 µg/mL and LOQ, 0.25 µg/mL. Thus, the proposed method proved to be easy, low cost, and accurate, and therefore, an useful alternative to quantify acyclovir in nanoparticles. |
| Formato |
text/html |
| Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000100035 |
| Idioma(s) |
en |
| Publicador |
Sociedade Brasileira de Química |
| Fonte |
Química Nova v.35 n.1 2012 |
| Palavras-Chave | #acyclovir #polymeric nanoparticles #derivative spectrophotometry |
| Tipo |