Avaliação da resistência à corrosão de implantes em aços inoxidáveis ISO 5832-9 e F138 removidos de pacientes


Autoria(s): Silva,Elison da Fonseca e; Oliveira,Luiz Fernando Cappa de
Data(s)

01/01/2012

Resumo

This work has compared the surfaces of two different steel samples used as orthopedical implants, classified as ASTM F138 and ISO5832-9, through optical emission spectroscopy, by means of SEM and EDS. The samples (implants) were also submitted to potentiodynamic cyclic polarization in Ringer lactate and NaCl 0.9 M L-1 solutions; ISO5832-9 sample did not show any kind of localized corrosion, but in the case of F138 steel was observed a pit localized corrosion in both solutions. In Ringer lactate solution it was observed a loss of about 63% for nickel and 26% for iron for F138 stell, compared to the initial composition.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000100012

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.35 n.1 2012

Palavras-Chave #stain steel #orthopedic implant #corrosion
Tipo

journal article