Validação de método para determinação de ácidos orgânicos voláteis em efluentes de reatores anaeróbios empregando cromatografia líquida


Autoria(s): Cerqueira,Maristela Barnes Rodrigues; Dias,Adriana Neves; Caldas,Sergiane Souza; Santana,Fabrício Butierres; D'Oca,Marcelo Gonçalves Montes; Primel,Ednei Gilberto
Data(s)

01/01/2011

Resumo

This work deals with the method validation for the determination of acetic, propionic and butyric acids (VFAs) in wastewaters from anaerobic reactors by HPLC-DAD. Separation was performed using a C18 column and the mobile phase composition were water pH 3.0 and methanol 90:10 (v/v). The detection and quantification was carried out at 220 nm. The method shows good linearity (r²>0.996), with adequate accuracy (89-102%) and relative standard deviations lower than 18%. The matrix effect was considered low (-4.1, -3.9 and 1.4%). The developed method is fast, simple and cheap; and it was applied in wastewater samples from anaerobic reactor.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422011000100029

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.34 n.1 2011

Palavras-Chave #VFAs #matrix effect #HPLC-DAD
Tipo

journal article