Determinação de 2,5-hexanodiona em urina empregando cromatografia líquida de alta eficiência, após derivatização com 2,4-dinitrofenil-hidrazina


Autoria(s): Antunes,Marina Venzon; Feltraco,Lilian de Lima; Morisso,Fernando Dal Pont; Linden,Rafael
Data(s)

01/01/2011

Resumo

A method for quantifying urinary 2,5-hexanedione was optimized and validated. Urine samples were hydrolyzed and derivatized with 2,4-dinitrophenylhydrazine. The analyte was separated in a high performance liquid chromatography system with a diode array detector, using a C18 column (150 x 4.6 mm, p.d. 5 µm) and a mobile phase composed of phosphate buffer pH 2.3:acetonitrile (40:60, v/v), at a flow rate of 1 mL/min. The chromatograms were monitored at 334 nm. Retention time was 7.3 minutes. Main validation parameters were: coefficient of determination: 0.9994, accuracy: 96 to 107%; intra-assay precision (RSD): 3.08 to 6.72%; inter-assay precision (RSD): 2.54 to 8.17% and limit of quantitation of 0.19 µg/mL.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422011000100028

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.34 n.1 2011

Palavras-Chave #2,5-hexanedione #HPLC #2,4-dinitrophenylhydrazine
Tipo

journal article